MARKETS
Laminar Field
Laminar flows ensures a local increase of the purity class in a cleanroom (Class "A" to GMP). Laminar field is used for cleanrooms that require not only clean air but also intensive air exchange. These laminar ceilings with an air flow and built-in HEPA filters consists of FFU units (circulation units) and are designed to be installed in suspending ceilings. Integrated LED lights are a part of the ceilings.
Description
- Designed for cleanrooms
- Use of the filter class H14
- Filters with a gel seal
- Local increase of purity class to qualification “A” according to GMP
- Flow speed 0,3-0,45 m.s-1 at air density r=1,2 kg/m3
- Automatically controlled flow speed
- Laminar field controlled by the Siemens Simatic system
- Possibility to install PVC lamella to direct the flow
- LED lighting 500lx
- Operating modes – Full-mode, muted mode and switched off
- Possibility to connect a speed sensor laminar flow
SIMILAR PRODUCTS